Show plain JSON{"containers": {"cna": {"affected": [{"defaultStatus": "unaffected", "product": "N'Vision Clinician Programmer", "vendor": "Medtronic", "versions": [{"status": "affected", "version": "all versions"}]}, {"defaultStatus": "unaffected", "product": "8870 N\u2019Vision removable Application Card", "vendor": "Medtronic", "versions": [{"status": "affected", "version": "all versions"}]}], "credits": [{"lang": "en", "type": "finder", "value": "Billy Rios and Jonathan Butts of Whitescope LLC reported these vulnerabilities to CISA."}], "datePublic": "2018-05-17T06:00:00.000Z", "descriptions": [{"lang": "en", "supportingMedia": [{"base64": false, "type": "text/html", "value": "<p>Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programme and 8870 N'Vision removable Application Card do not encrypt PII and PHI while at rest.</p>"}], "value": "Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programme and 8870 N'Vision removable Application Card do not encrypt PII and PHI while at rest."}], "metrics": [{"cvssV3_1": {"attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "NONE", "baseScore": 4.6, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1"}, "format": "CVSS", "scenarios": [{"lang": "en", "value": "GENERAL"}]}], "problemTypes": [{"descriptions": [{"cweId": "CWE-311", "description": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311", "lang": "en", "type": "CWE"}]}], "providerMetadata": {"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert", "dateUpdated": "2025-06-27T16:24:54.910Z"}, "references": [{"name": "104213", "tags": ["vdb-entry", "x_refsource_BID"], "url": "http://www.securityfocus.com/bid/104213"}, {"tags": ["x_refsource_MISC"], "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-137-01"}, {"tags": ["x_refsource_CONFIRM"], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf"}, {"url": "https://www.medtronic.com/security"}], "source": {"advisory": "ICSMA-18-137-01", "discovery": "EXTERNAL"}, "title": "Medtronic N'Vision Clinician Programmer Missing Encryption of Sensitive Data", "workarounds": [{"lang": "en", "supportingMedia": [{"base64": false, "type": "text/html", "value": "<p>Medtronic has not developed a product update to address the \nvulnerabilities, but is reinforcing security reminders within this \nadvisory to help reduce the risk associated with the vulnerabilities.</p>\n<p>The 8870 Therapy Application card stores PHI and PII as part of its \nnormal operating procedure and should be handled, managed and secured in\n a manner consistent with the applicable laws for patient data privacy.</p>\n<p>Medtronic recommends users take additional defensive measures to \nminimize the risk of exploitation of these vulnerabilities. \nSpecifically, hospitals and clinicians should:</p>\n<ul>\n<li>Maintain strict physical control of the 8870 application card.</li>\n<li>Use only legitimately obtained 8870 cards and not cards provided by \nany third party as firmware and system updates are provided directly by \nMedtronic using new 8870 application cards.</li>\n<li>8840 Programmers and 8870 Therapy Application compact flash cards \nare the property of Medtronic and should be returned to Medtronic when \nno longer in use. If that is not an option, you should securely dispose \nof them.</li>\n</ul>\n<p>Medtronic has released additional patient focused information, at the following location:</p>\n<p><a target=\"_blank\" rel=\"nofollow\" href=\"https://www.medtronic.com/security\">https://www.medtronic.com/security</a></p>\n\n<br>"}], "value": "Medtronic has not developed a product update to address the \nvulnerabilities, but is reinforcing security reminders within this \nadvisory to help reduce the risk associated with the vulnerabilities.\n\n\nThe 8870 Therapy Application card stores PHI and PII as part of its \nnormal operating procedure and should be handled, managed and secured in\n a manner consistent with the applicable laws for patient data privacy.\n\n\nMedtronic recommends users take additional defensive measures to \nminimize the risk of exploitation of these vulnerabilities. \nSpecifically, hospitals and clinicians should:\n\n\n\n * Maintain strict physical control of the 8870 application card.\n\n * Use only legitimately obtained 8870 cards and not cards provided by \nany third party as firmware and system updates are provided directly by \nMedtronic using new 8870 application cards.\n\n * 8840 Programmers and 8870 Therapy Application compact flash cards \nare the property of Medtronic and should be returned to Medtronic when \nno longer in use. If that is not an option, you should securely dispose \nof them.\n\n\n\n\nMedtronic has released additional patient focused information, at the following location:\n\n\n https://www.medtronic.com/security"}], "x_generator": {"engine": "Vulnogram 0.2.0"}, "x_legacyV4Record": {"CVE_data_meta": {"ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2018-05-17T00:00:00", "ID": "CVE-2018-8849", "STATE": "PUBLIC"}, "affects": {"vendor": {"vendor_data": [{"product": {"product_data": [{"product_name": "N'Vision Clinician Programmer", "version": {"version_data": [{"version_value": "8840 N'Vision Clinician Programmer, all versions"}, {"version_value": "8870 N'Vision removable Application Card, all versions"}]}}]}, "vendor_name": "Medtronic"}]}}, "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": {"description_data": [{"lang": "eng", "value": "Medtronic N'Vision Clinician Programmer 8840 N'Vision Clinician Programmer, all versions, and 8870 N'Vision removable Application Card, all versions does not encrypt PII and PHI while at rest."}]}, "problemtype": {"problemtype_data": [{"description": [{"lang": "eng", "value": "MISSING ENCRYPTION OF SENSITIVE DATA CWE-311"}]}]}, "references": {"reference_data": [{"name": "104213", "refsource": "BID", "url": "http://www.securityfocus.com/bid/104213"}, {"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01", "refsource": "MISC", "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"}, {"name": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf", "refsource": "CONFIRM", "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf"}]}}}, "adp": [{"providerMetadata": {"orgId": "af854a3a-2127-422b-91ae-364da2661108", "shortName": "CVE", "dateUpdated": "2024-08-05T07:10:45.899Z"}, "title": "CVE Program Container", "references": [{"name": "104213", "tags": ["vdb-entry", "x_refsource_BID", "x_transferred"], "url": "http://www.securityfocus.com/bid/104213"}, {"tags": ["x_refsource_MISC", "x_transferred"], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-18-137-01"}, {"tags": ["x_refsource_CONFIRM", "x_transferred"], "url": "http://www.medtronic.com/content/dam/medtronic-com/us-en/corporate/documents/Medtronic-NVision-8840_Security-Bulletin_FINAL.pdf"}]}]}, "cveMetadata": {"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "assignerShortName": "icscert", "cveId": "CVE-2018-8849", "datePublished": "2018-05-18T13:00:00Z", "dateReserved": "2018-03-20T00:00:00", "dateUpdated": "2025-06-27T16:24:54.910Z", "state": "PUBLISHED"}, "dataType": "CVE_RECORD", "dataVersion": "5.1"}